Myers Vacuum
REPAIR SERVICES, INC

GTC-390 Thermocouple
Dual Bar Graph

GTC-390 Product Manual
click to download

Note: Requires the Adobe Acrobat Reader.


Features

  • Dual pressure display
  • Measures from of 1 m torr to 5000 m torr
  • Each station has a relay output adjustable from the front panel
  • Front panel LED for relay output status
  • Panel mountable or beach top use
  • Uses the rugged and proven GTC-036 senor

Description
The GTC-390 thermocouple vacuum gauge is a dual station thermal conductivity type gauge and controller housed in a DIN-style case. The unit uses the rugged and proven GTC-036 sensor and covers a pressure range of 1 mtorr to 5000 mtorr. Pressures are displayed simultaneously for both stations on two 40-segment bar graph displays. Each station includes a relay set point that can be set within the range of the gauge. The relay set point for each station may be viewed by pressing the “View” buttons on the front panel. Adjusting the set points is accomplished by turning the front panel potentiometers marked “Set”. An LED indicator displays the output status of each relay control output.


Principal of Operation
The GTC-390 thermocouple gauge consists of two identical pressure measurement and control circuits that function as follows: A constant current source is supplied to the heating element of the GTC-036 sensor. The thermocouple junction in the sensor generates a voltage signal that varies with the temperature of this heating element. A decrease in pressure reduces the number of gas molecules available to transfer heat away from the heating element, resulting in a temperature rise and an increased output from the sensor. This small voltage signal is filtered, amplified, and inverted resulting in a signal with a span of several volts that is proportional to pressure. This pressure signal is then passed through a signal selection circuit to the display drivers where it is compared to a high and a low reference and the proper display segments are illuminated. A voltage comparator is used to activate the set point relay as well as the front panel set point indicator. There is a small amount of hysteresis built into this circuit. The “View” button causes the signal selection circuit to send the pressure set point level to the display drivers rather than the pressure.

Specifications

GAUGE SPECIFICATIONS GTC-390
Power Input 100-120 VAC 50/60 Hz 40 watts
or 9 VDC 9 watts
Pressure Scales 1 mtorr to 5000 mtorr on a 40-segment bar-graph
Number of Sensing Stations 1 or 2
Pressure Sensing Tube Type GTC-036 with 1/8” pipe thread or 0.406” diameter O-Ring sealing surface.
Tube Cable 10 ft length standard with both 9-pin gauge and octal sensor connector
Cabinet Dimensions 5.67” wide 5.10” deep
2.84” high
Mounting Hole Dimensions 5.45” wide x 2.68” high (allow 8” for mounting depth)
Relay Connections One removable three position terminal block for each station with common, normally closed, normally open connections
Relay Trip Point Adjustment 1 mtorr to 5000 mtorr
Relay Types SPDT 5 Amp 250 VAC Max

GTC-036 Tube Specifications
http://www.myers-vacuum.com/gtc036.shtml

Ordering information for the components mentioned on this page is listed below.

ITEM
DESCRIPTION
P/N
GTC-390 THERMOCOUPLE din panel mount, dual 40 segment bar graph display, with single set point, pressure scale of 0-5 torr with cables, less sensors 500623-0210
GTC-390 As above - complete with two GTC036 sensors 500623-0212

To change the combination for ordering the first dash number is always –0 the second dash indicates the number of station cable (zero, one or two), the third dash number determines the length of the connector cable, -0010 = 10’ connector cable, -0020 = 15’ connector cable, -0030 =25’ connector cable, -0040 =50’ connector cable and the fourth dash number indicates the number of sensors (zero, one or two).

 