| GT-430
Thermistor
Digital Display with Trend SensingTM technology
The
430 THERMISTOR GAUGE reads from atmosphere to
1 X 10-3 Torr plus Trend SensingTM,
a unique feature that allows the user to see which direction the
pressure is going.
- Trend
SensingTM shows direction of pressure change
- Easy to
read pressure display
- Measures
from atmosphere to 1 X 10-3 torr
- Two adjustable
independent relay outputs
- LED indication
of Pressure range, set point indication/set value and status
- RS-232
serial interface
- Analogue output
(0-3 volts) for remote display/monitoring
- Leak detection
function over vacuum range
- Front panel
LED’s for relay output status
- Panel mountable
(1/8 DIN) or bench top use
- Uses GT-034
sensor
The GT-430
thermistor vacuum gauge is a digital single station gauge for
use in harsh environments. It incorporates Trend SensingTM
technology that indicates which direction pressure is going. Trend
SensingTM enhances vacuum leak detection capabilities.
The controller is housed in a 1/8 DIN-style case. The unit uses
the rugged and proven GT-034 thermistor sensor and covers a pressure
range of 1 X 10-3 Torr to atmosphere. Pressure is displayed
4 digit monitor with torr and millitorr indicators. A nine segmented
bar graph displays pressure direction. This segmented display
provides high sensitivity that it can be used for vacuum leak
detection with a solvent. The relay set point for each station
may be viewed by pressing the “View” buttons on the
front panel. The gauge features two independent relays that can
be set within the range of the gauge, each set point includes
a separate high and low pressure value. Three front panel buttons
allow the user to toggle through and adjust the values. LED’s
indicate which value is being displayed with separate LED’s
to indicate the status of the set points. The unit features an
RS-232 serial interface that allows a remote system to request
information and a separate linear analog voltage output that is
proportional to pressure.
A single
resistor element is employed in a bridge-type detector driven
from a constant regulated power source. The inverse resistive
characteristics of the sensor unbalances the bridge as pressure
changes occur, thereby causing a current change. A microprocessor
scales the changes into pressure units. Trend SensingTM
electronics enhances the sensitivity of each range by displaying
the rate of change of the pressure on a separate bar graph.
Trend
SensingTM technology puts the gauge in constant
Leak Detect Mode. Front panel Bar graph high sensitivity indicates
to vacuum pressure changes. The bar graph will give quick feedback
when the pressure makes a rapid change.
SPECIFICATIONS |
| ITEM
|
DESCRIPTION
|
Power
Input |
100-120
VAC 50/60 Hz 15 watts or 9-12 VDC |
Pressures
Scales |
760
Torr to 1 X 10-3 Torr |
Number
of Sensing Stations |
One |
Pressure
Sensing Tube |
Type
GT-034 with 1/8”
pipe thread or 0.406” diameter O-Ring sealing surface.
|
Tube
Cable |
10
ft length standard with both 9-pin gauge and octal sensor
connector |
Cabinet
Dimensions |
3.78”
wide 5.1” deep 1.89” high |
Mounting
Hole Dimensions |
3.64"
wide x 1.75” high (allow 8” for mounting depth)
|
Relay
Connections |
One
removable six position terminal block with common, normally
closed, normally open connections for each set point. |
Relay Trip Points Adjustments |
Entire range of gauge |
Relay
Types |
SPDT 5 Amp 250 VAC Max |
RS-232
|
RX,
TX, and COM (9600 Baud) |
Analog
Output |
0-3 Volt, (COM and OUT) |
GT-034
Tube Specifications http://www.myers-vacuum.com/gt034.shtml
ORDERING
INFORMATION |
ITEM |
DESCRIPTION |
P/N
|
GT-430 |
THERMISTOR
GAUGE - DIN panel mount with digital display dual set points
scale indication from atmosphere to 0 torr with cable, less
sensor |
500681-0010 |
GT-430 |
As above - complete with GT034 sensor |
500681-0011 |
To change the combination
for ordering the first dash number is always –0 the second
dash is always-00, the third dash number determines the connector
cable and its length ,-0000= no cable -0010 = 10’ connector
cable, -0020 = 15’ connector cable, -0030 =25’ connector
cable, -0040 =50’ connector cable and the fourth dash number
indicates with or without sensor (zero or one).
|